Support

Troubleshooting Guide

1. Can’t login
  1. Upon sign-up an email will be sent to the provided email, this must be confirmed before the account is active
  2. Reset your password by using the Forgot Password button on the login screen
2. Not all app features are available

New users joining a company must be approved by a company or group admin before all the app features become available; admins can do this either by accepting the email sent or through their notifications in the app

3. I cannot connect the app to the ROAM instrument
  1. Verify smart device’s Bluetooth is turned on
  2. Reset the ROAM instrument by holding down the button on the device itself
  3. Verify that the app is up to date
  4. Close the app and then re-open
  5. Delete and then re-install the app
4. I’m not seeing all the parameters I want to test
  1. Ensure that parameters are turned on by going to Settings. Please note that turning off Free Chlorine, Total Chlorine, and Chloramines, or alkalinity will turn off associated parameters as well
  2. Check that desired parameters are active in Settings > Parameters, deactivated parameters are not displayed but are still collected and will appear by reactivating the parameter
5. Light turns red when trying to start a test
  1. Using the app connect to the ROAM to check if the cartridge is expired or has used all tests
  2. Remove the cartridge from the ROAM device, clean the connector then plug it back in
  3. Reset the ROAM instrument by holding the button on the device itself
6. Calibration is not passing
  1. Verify the cartridge is fully plugged into the ROAM
  2. Remove the cartridge from the ROAM device, clean the connector then plug it back in
  3. Help proper mating by holding the cartridge in tight during the first 30 seconds of calibration
  4. Reset the ROAM instrument by holding the button on the device itself
7. Results seem inaccurate
  1. Verify the cartridge is fully plugged into the ROAM
  2. Remove the cartridge from the ROAM device, clean the connector then plug it back in
  3. Recalibrate the cartridge by connecting to it through the app
8. Chlorine is reporting lower than expected or there is more spread between Free and Total Chlorine than expected
  1. Bleach and rinse the vials
  2. Rinse and clean the cartridge pockets where the sample and waste vials sit
9. Nothing happens when the button is pressed
  1. Plug ROAM into USB charger, if the charging indicator turns on please allow time to charge before running a test, if fully drained it may take a few minutes for the charging indicator to turn on
  2. Unplug the cartridge and reset the ROAM by holding the button
10. Unresolved issues

For any unresolved issues please contact our support team at 801-839-1071

Need More Help? Contact Us

Any questions? Please let us know.